Effectiveness and user experience of nose and throat swabbing techniques for SARS-CoV-2 detection: results from the UK COVID-19 National Testing Programme

Abstract

Background

The UK’s National Health Service Test and Trace (NHSTT) program aimed to provide the most effective and accessible SARS-CoV-2 testing approach possible. Early user feedback indicated that there were accessibility issues associated with throat swabbing. We report the results of service evaluations performed by NHSTT to assess the effectiveness and user acceptance of swabbing approaches, as well as qualitative findings of user experiences from research reports, surveys, and incident reports. Our intent is to present and summarize our findings about the application of alternative swabbing approaches during the COVID-19 pandemic in the UK.

Methods

From May 2020 to December 2021, NHSTT conducted a series of service evaluations assessing self-swabbing and assisted swabbing of the nose and throat, and nose only (anterior nares/mid-turbinate) using polymerase chain reaction (PCR) and lateral flow devices (LFDs), for diagnostic suitability within the COVID-19 National Testing Programme. Outcomes included observational user feedback on swabbing approaches and quantitative testing performance (concordance, sensitivity, and specificity). A post-hoc indirect comparison of swabbing approaches was also performed. Additionally, an analysis of existing cross-service research was conducted in April 2021 to determine user feedback regarding swabbing approaches.

Results

Observational data from cross-service research indicated a user preference for nose swabbing over throat swabbing. Significantly more users reported that nose swabbing was easier to perform than throat swabbing (50% vs. 12%) and there were significantly fewer reported incidents. In the service evaluations, while there was reduced sensitivity for nose-only swabbing for PCR (88%) compared with nose and throat swabbing, similar sensitivities were observed for nose-only and nose and throat swabbing for LFDs. The sensitivity of nose-only swabbing for LFDs was higher for individuals with higher viral concentrations.

Conclusions

User experience analyses supported a preference for nose-only swabbing. Nose-only swabbing for LFDs provided sufficient diagnostic accuracy, supporting its use as a viable option in the COVID-19 National Testing Programme. Less invasive swabbing approaches are important to maximize testing accessibility and alongside other behavioral interventions, increase user uptake.

Peer Review reports

Background

In response to the SARS-CoV-2 pandemic, the UK Government established the COVID-19 National Testing Programme and committed to mass testing, with initial testing commencing in March 2020 [1,2,3]. By May 2020, the program had evolved into the National Health Service Test and Trace (NHSTT) service with the objective to provide the most effective testing approach on a national scale [1, 4]. Accessibility of testing was one of the main focuses of the program and regional or local testing sites were rapidly established to allow SARS-CoV-2 testing [5]. Other initiatives to increase testing accessibility included exploring potentially easier sample collection approaches, such as the viability of saliva as a sampling method and whether remote support for testing could improve access to people who are blind and partially sighted [6].

In the initial period mainly molecular tests like quantitative reverse-transcription polymerase chain reaction (qRT-PCR, referred to as PCR hereafter), a highly sensitive and specific test that uses nucleic acid amplification to detect viral RNA [7,8,9,10,11,12] were used. While effective, PCR testing capacity was initially limited by the need for trained staff to undertake swabbing, a finite number of appointment slots, the need for individuals to attend an in-person testing site, high cost per test, and long turnaround times (24–72 h due to capacity). These limitations were overcome by the introduction of rapid antigen testing, which uses a point-of-care lateral flow device (LFD) to detect viral surface proteins produced during the active phase of SARS-CoV-2 infection [10]. Although not as sensitive as PCR, LFDs have several advantages as the results are usually available within 15–30 min and their low cost and convenience allow for a much higher volume of testing and broader accessibility [10, 13,14,15,16]. LFD tests were swiftly developed and evaluated for use in the UK by the late summer/autumn of 2020 [17,18,19,20]. PCRs were initially used in hospitals for the identification of cases, to prevent incursions of SARS-CoV-2 into healthcare settings, and for infection prevention and control in clinical settings. Outside of their use in clinical care, PCR testing was predominantly for testing symptomatic individuals for public health purposes (i.e., to identify if self-isolation and contact tracing were required). It was also used for asymptomatic testing with weekly PCR tests supplemented by further LFD testing. LFDs were initially used for asymptomatic mass testing and in research (SARS-CoV2 immunity and reinfection evaluation (SIREN) study) [21]. Over the course of the pandemic, LFD testing was expanded to symptomatic individuals to help people manage risk in assessing infectiousness to guide self-isolation timelines [22].

In the UK, performance requirements for PCR testing as part of the National COVID-19 Testing Programme were based on meeting the same standards as for clinical care [12]. For LFDs requirements for inclusion in the UK program were based on meeting laboratory performance evaluation criteria [17]. While the real-world performance for the first regulatory-approved self-test LFD [23] was generally considered a minimum benchmark, there was ongoing debate about the level of test performance that would be required to have a positive public health impact [24,25,26]. Specifically, questions arose regarding the importance of detecting infectious individuals (those with high viral concentrations) versus simply identifying infection [24, 26,27,28]. This created an urgent need to understand levels of performance that could be achieved to consider how LFDs may be effectively implemented.

The UK’s testing strategy and resultant policies were adapted and revised frequently by policymakers based on the changing epidemiology of the SARS-CoV-2 pandemic, updated scientific evidence, and the technology available at the time as well as socio- and economic factors [29,30,31,32]. As new diagnostic testing strategies became available and preferences for swabbing became apparent, NHSTT (with the Department of Health and Social Care [DHSC], later becoming part of the UK Health Security Agency) performed a series of service evaluations to assess the effectiveness and user acceptance of new testing and swabbing approaches, as well as to capture user experience and feedback. By analyzing both quantitative data from service evaluations and qualitative user feedback, we aim to provide insights into the most effective and accessible testing methods. The findings from this study are particularly relevant for informing testing strategies in future pandemics, where rapid and widespread testing will again be crucial.

Methods

NHSTT user attitude survey and analysis

The NHSTT LFD Product Research Team performed a post-hoc analysis to compare user preferences for nose and throat swabbing versus nose-only swabbing using existing cross-service user research as of April 2021. The existing research included testing surveys from service users and testing site leads, cross-service evaluation reports, and incident management reports related to swabbing. The goal was to see if nose-only swabbing would increase testing uptake. Data was analyzed in R (version 4.3.2) using prop.test, which implements a chi-squared test for equality of proportions with Yate’s continuity correction, to assess swab preferences and swabbing-related incidents.

Between March 4, 2021, and February 14, 2023, the Voice of the Customer (VOTC) program [16] captured feedback from SARS-CoV-2 testing experiences in England. Surveys were emailed to participants and data were collected regarding key touchpoints with the service: test, trace, and isolation. Responses were analyzed for common detractors, and the data was used to create a word cloud. Detractor proportions were calculated with R (version 4.3.2) using prop.test as above.

Service evaluations and statistical analysis

The service evaluations reported here were part of the wider program conducted in the UK from May 2020 to December 2021 Fig. 1. Service evaluations compared swabbing techniques and testing performances, including if self-swabbing (SS) with LFDs at home was as effective as testing at a testing site. User feedback on swabbing experiences, where participants rated discomfort or pain during the process was collected as part of observational data. An overview of the service evaluations including start date, testing method, and swabbing technique can be found in Table 1.

Fig. 1
figure 1

Timeline of key UK service evaluations. Figure is for illustration purposes, where the boxes are reflective of approximate study start date. NT-PCR1 to NT-PCR4: AS versus SS of the nose and throat using PCR. SS-AN-PCR1: SS of the anterior nares using PCR. AS-NT-LFD1: AS of the nose and throat using the Innova LFD. SS-NT-LFD1, SS-NT-LFD2: SS of the nose and throat using the Innova LFD. AS-AN-LFD1: AS of nose and throat and anterior nares using the Orient Gene LFD. SS-AN-LFD1: SS of anterior nares using the Orient Gene LFD. SS-AN-LFD2: SS of anterior nares using the SureScreen LFD. SS-NT-LFD3: SS of the nose and throat using the Innova LFD at participants’ home. SS-MT-LFD1: SS of both nostrils to mid-turbinate level using the Innova LFD. SS-AN-LFD3: SS of anterior nares using the SureScreen LFD at participants’ home. SS-AN-LFD4: SS of anterior nares using the Innova LFD at participants’ home. AS-AN-LFD2: AS of anterior nares using Innova LFD. AN, anterior nares; AS, assisted swabbing; LFD, lateral flow device; MT, mid-turbinate level, NT, nose and throat; NHSTT, National Health Service Test and Trace; PCR, polymerase chain reaction; SS, self-swabbing; VOTC, Voice of the Customer

Table 1 Overview of the service evaluations

Quantitative data was collected from various evaluations to compare the performance of the devices and swabbing approaches. Participants either swabbed their throat and nose or performed nose-only swabbing. For LFD tests, participants read and reported results themselves, while PCR samples were sent to laboratories. Sensitivity and specificity of LFDs were compared with PCR results, no specific performance benchmarks were set prior to the evaluations.

Matched LFD and qRT-PCR results were analyzed to determine concordance, sensitivity (stratified by viral concentration), and specificity. To calculate the 95% confidence intervals (CI) for sensitivity and specificity estimates, and observational data, the Clopper-Pearson exact method was used. Post-hoc analyses assessing the differences of proportions in the observational data were calculated in R (version 4.3.2) using prop.test.

Methodology and device performance data for some of these service evaluations have been partially reported elsewhere as part of the UK DHSC’s commitment to evaluating the performance of in-vitro diagnostic devices for use in the COVID-19 National Testing Programme [18, 20, 33,34,35,36].

Methods including specifics regarding the LFDs and PCR tests conducted in the service evaluations are detailed in the Additional file 1: Supplementary methods.

Results

User feedback on swabbing and testing approaches

NHSTT analysis of user preferences with LFD testing

From the testing surveys assessing swab preferences of the UK general public (N = 54,415) for PCR tests, significantly more participants found nose swabs easier to use compared with throat swabs (50% vs. 12%; p < 0.0001) (Additional file 1: Fig. S1A). From the cross-survey research reports, despite the small total number of reported incidents (N = 71), significantly more incidents of broken swabs were recorded through incident reporting with throat swabbing than with nose swabbing (p < 0.0001; Additional file 1: Fig. S1B).

VOTC user attitudes survey

Between March 4, 2021, and February 14, 2023, 1,289,611 entries were recorded in the VOTC survey regarding attitudes towards SARS-CoV-2 testing. Of these, 46,603 included the description of detractors of testing with 2473 entries including the word “swab” (Additional file 1: Fig. S2). Among these 29% included “throat” versus 16% included “nose”. The difference between the proportions was highly statistically significant (p < 0.0001) indicating a stronger negative association of throat swabbing compared with nose swabbing.

Observational data from the service evaluations

A total of 1392 participants from studies NT-PCR1 to NT-PCR4 provided feedback on their experience with nose and throat swabbing. Significantly greater proportions of participants found throat swabbing to be more uncomfortable than nose swabbing (63% vs. 43%), and nose swabbing to be completely painless compared with throat swabbing (21% vs. 14%) (p < 0.0001; Fig. 2A). In study SS-AN-PCR1 (n = 1962), similar proportions of participants found nose swabbing (anterior nares, swallow nose) and throat swabbing to be uncomfortable (52% vs. 51%), but significantly greater proportions of participants found nose swabbing to be completely painless (33% vs. 23%; p < 0.0001) (Fig. 2B). In contrast, significantly greater proportions of patients reported that mid-turbinate (deeper nose) swabbing was uncomfortable compared with nose and throat swabbing (48% vs. 34%; p < 0.0001) in study SS-MT-LFD1 (n = 1148) (Fig. 2C).

Fig. 2
figure 2

Observational user feedback and insight on experiences with swabbing approaches from seven UK service evaluation studies. A NT-PCR1 to NT-PCR4 (SS and AS of nose and throat). B SS-AN-PCR1 (SS anterior nares). C SS-MT-LFD1 (SS of both nostrils to mid-turbinate level). Data are mean and 95% CIs. p values derived from chi-sq tests comparing nose vs. throat swabbing. Blue bars represent nose-only swabbing (including nose-only and the nose part of nose and throat swabbing) for PCR. Yellow bars represent throat-only swabbing (the throat part of nose and throat swabbing) for PCR. Navy bars represent mid-turbinate nose swabbing for LFDs. Orange bars represent nose and throat swabbing for PCR. A NT-PCR1 to NT-PCR4: AS vs. SS of the nose and throat for PCR. B SS-AN-PCR1: SS of the anterior nares for PCR vs. SS of nose and throat for PCR. C SS-MT-LFD1: SS of both nostrils to mid-turbinate level using the Innova LFD vs. SS of the nose and throat for PCR. *p < 0.005; **p < 0.0005; ***p < 0.0001. AN, anterior nares; AS, assisted swabbing; CI, confidence interval; LFD, lateral flow device; MT, mid-turbinate, NT, nose and throat; PCR, polymerase chain reaction; SS, self-swabbing

Service evaluations: quantitative testing outcomes

An overview of the number of participants recruited, the paired samples evaluable for analysis, and the quantitative testing outcomes in the service evaluations are shown in Table 1. Numbers of void samples and other omitted samples are shown in Additional file 1: Table S2. Key findings of the service evaluations and subsequent indirect comparison are reported below according to the different research questions covered in this report.

Can shallow nose swabbing provide the same performance as nose and throat swabbing using PCR tests?

Study SS-AN-PCR1 evaluated the effectiveness of self-swabbing (SS) of the anterior nares (shallow nose) for identifying SARS-CoV-2 compared with SS of the nose and throat for PCR. Of the 2083 paired PCR samples evaluable, 97.1% were concordant for positive (n = 336) or negative (n = 1688) results. Of the 59 discordant samples, 46 were positive in the nose and throat samples only (Additional file 1: Table S3). There was a statistically significant difference between discordant results in favor of nose and throat swabbing versus nose-only swabbing (p < 0.001). The overall sensitivity of SS of the anterior nares for PCR was 88% (Table 1; Fig. 3).

Fig. 3
figure 3

Overall point sensitivities of key UK service evaluations. SS-AN-PCR1: SS of the anterior nares using PCR. SS-AN-LFD1: SS of anterior nares using the Orient Gene LFD. SS-AN-LFD2: SS of anterior nares using the SureScreen LFD. SS-AN-LFD3: SS of anterior nares using the SureScreen LFD at participants’ home. SS-AN-LFD4: SS of anterior nares using the Innova LFD at participants’ home. SS-NT-LFD1, SS-NT-LFD2: SS of the nose and throat using the Innova LFD. SS-NT-LFD3: SS of the nose and throat using the Innova LFD at participants’ home. AS-NT-LFD1: AS of the nose and throat using the Innova LFD. AS-AN-LFD1: AS of nose and throat and anterior nares using the Orient Gene LFD. SS-MT-LFD1: SS of both nostrils to mid-turbinate level using the Innova LFD. AS-AN-LFD2: AS of anterior nares using Innova LFD. Sensitivity = TP/(TP + FN) based on corresponding qRT-PCR test results; binomial 95% CIs, Pearson-Klopper method. AN, anterior nares; AS, assisted swabbing; CI, confidence interval; FN, false negative; LFD, lateral flow device; MT, mid-turbinate level, NT, nose and throat; PCR, polymerase chain reaction; SS, self-swabbing; TP, true positive. The dotted vertical line indicates a diagnostic sensitivity of 50%, which was deemed sufficient in a real-world setting for use in the NHSTT program

Analysis of 109 paired samples classified into differing viral concentration groups based on PCR cycle threshold value (Ct), 87 showed higher concentrations for nose and throat swabbing and 38 showed higher concentrations for nose-only swabbing (p < 0.001) (Additional file 1: Table S4). Of the 46 samples which were positive for nose and throat swabbing and negative for nose-only swabbing, 18 had a Ct value of < 25 (higher viral concentration > 10,000 (10 K) viral copies/mL).

The average Ct score for nose and throat swabbing was 21.6 (standard error [SEM] 0.24) and for nose-only swabbing was 22.6 (SEM 0.28), indicating a lower viral concentration with anterior nares swabbing (p < 0.001; t-test). There was a strong correlation (Spearman correlation r = 0.87, p < 0.001) between the average Ct scores of the nose and throat, and nose-only samples (Fig. 4).

Fig. 4
figure 4

Comparison of SARS-CoV-2 viral concentration in SS-AN-PCR1. A Average Ct of nose and throat versus nose only swabs in SARS-CoV-2–positive individuals. B Correlation of Ct between nose and throat versus nose only swabs in all individuals. SS-AN-PCR1: SS of the anterior nares using PCR. A The average Ct value was taken across the three genes (ORF1ab gene, N gene, and S-gene) in all concordant and discordant SARS-CoV-2–positive individuals; the boxplot shows a comparison between the SS and AS samples. There was a statistically significant difference in Ct values for nose and throat swabs versus nose-only swabs: mean − 0.99; 95% CI − 1.3, − 0.6; p < 0.001. B There was a strong correlation (Spearman’s rho correlation co-efficient = 0.87; p < 0.001) between the average Ct values of the nose and throat, and nose-only samples. AN, anterior nares; CI, confidence interval; Ct, cycle threshold; NT, nose and throat; PCR, polymerase chain reaction; SS, self-swabbing

Overall, SS-AN-PCR1 demonstrated that SS of anterior nares has reduced sensitivity (88%) versus nose and throat swabbing when used for PCR testing as less viral material tended to be collected.

Can nose swabbing (via anterior nares or mid-turbinate level) provide the same diagnostic performance as nose and throat swabbing using LFDs?

Assisted swabbing of anterior nares

Study AS-AN-LFD1 assessed whether assisted swabbing (AS) of both nose and throat and double anterior nares techniques using the Orient Gene COVID-19 Ag Rapid Test Cassette provided sufficient diagnostic performance. For the phase of the study assessing swabbing of the anterior nares, there were 2597 evaluable paired PCR and LFD samples, of which 92.8% were concordant (n = 207 positive and n = 2204 negative) (Additional file 1: Table S5). Overall sensitivity for anterior nares swabbing was 53.2%, with a higher sensitivity of 88.3% for samples with high viral concentration (> 1 million (1 M) viral copies/mL). Overall sensitivity was 53.2%, with a higher sensitivity of 88.3% for samples with high viral concentration (> 1 M copies/mL), (Additional file 1: Table S6) [20]. Assisted swabbing of the nose and throat had a non-statistically higher sensitivity (59.6%) compared to double anterior nares (53.2%) (Table 1, Fig. 3, Additional file 1: Table S6).

Study AS-AN-LFD2 investigated the effectiveness and suitability of AS of both anterior nares using the Innova LFD test. Of the 1160 evaluable paired PCR and LFD tests, 89.2% were concordant for positive (n = 163) or negative results (n = 872) (Additional file 1: Table S5). Overall sensitivity was 57.6% with a higher sensitivity of 89.5% for samples with high viral concentration (> 1 M copies/mL; Additional file 1: Table S6). Comparison of overall sensitivity in AS-AN-LFD2 (anterior nares) and AS-NT-LFD1 (nose and throat [35]) suggested a similar sensitivity overall (57.6% vs. 54.4%) with overlapping 95% CIs indicating similar performance depending on viral concentration (Table 1, Fig. 3, Additional file 1: Table S6).

Self-swabbing of anterior nares and/or mid-turbinate level

Study SS-MT-LFD1 evaluated the effectiveness of SS of both nostrils to mid-turbinate level (deeper nose) using the Innova LFD test. A total of 91.8% of participants with paired PCR and LFD samples were concordant for a positive (n = 237) or negative (n = 775) result (Additional file 1: Table S7). Overall sensitivity for LFD double nose (to mid-turbinate level) swabbing was 72.9% and was higher (98.1%) for samples with viral concentration > 1 M copies/mL (Additional file 1: Table S8) [34]. Indirect comparison of overall sensitivity observed in studies SS-MT-LFD1 and SS-NT-LFD2 [34], suggested significantly greater sensitivity for SS-MT-LFD1 compared to SS-NT-LFD2 (Table 2).

Table 2 Indirect comparison of sensitivity of different self-swabbing approaches (mid-turbinate vs. nose and throat)

Two further studies, SS-AN-LFD1 (Orient Gene COVID-19 Ag Rapid Test Cassette) and SS-AN-LFD2 (SureScreen LFD) assessed the effectiveness of SS of the anterior nares using LFDs (Table 1, Fig. 3, Additional file 1: Tables S7 and S8). For the study, SS-AN-LFD1, 89.4% of samples were concordant for positive (n = 204) or negative (n = 1395) results. Sensitivity was 53.3% overall, and 87.8% for samples with viral concentration > 1 M copies/mL [20]. SS-AN-LFD2 had similar results with the SureScreen LFD, reporting an overall sensitivity of 71.0%, which was higher (95.2%) for samples with viral concentration > 1 M copies/mL.

These studies demonstrated that SS and AS of the anterior nares and SS of both nostrils to mid-turbinate level for LFDs provide sufficient diagnostic sensitivity which was comparable to nose and throat swabbing for the same LFD test. SS of both nostrils to mid-turbinate level had greater sensitivity than SS of the nose and throat for the Innova LFD.

Can self-swabbing and sampling at users’ homes provide the same diagnostic accuracy as self-testing at a dedicated testing site?

Self-swabbing of the nose and throat

Study SS-NT-LFD3 explored whether participant-reported performance using the Innova LFD (via SS of nose and throat at home) provided sufficient diagnostic accuracy and was comparable to that at a test site. Of the 773 evaluable paired samples, 46 were concordant for a positive result and 690 for a negative result (Additional file 1: Table S9). Overall sensitivity and specificity of the Innova LFD at home were 56.8% and 99.7%, respectively (Table 1). Sensitivity was higher (90.6%) for viral concentrations > 1 M copies/mL (Additional file 1: Table S10). The performance of the Innova LFD when used at home was similar to that in SS-NT-LFD1 [33, 35] (Table 1, Fig. 3, Additional file 1: Table S7) when the sample was taken at a dedicated test site, based on overlapping 95% CIs.

Self-swabbing of the anterior nares

Studies SS-AN-LFD3 and SS-AN-LFD4 assessed the effectiveness of SS of the anterior nares with the SureScreen LFD and Innova LFD, respectively, in a home setting. A total of 93.6% paired PCR and LFD samples were concordant in SS-AN-LFD3 (Additional file 1: Table S9). Sensitivity was 74.7% overall (Table 1, Fig. 3) and 93.3% for samples with viral concentration > 1 M copies/mL (Additional file 1: Table S10). Comparison of the sensitivity of the SureScreen LFD test in the SS-AN-LFD3 (home setting) and SS-AN-LFD2 (test site) studies suggested a similar sensitivity overall with overlapping 95% CIs (Table 1, Fig. 3).

SS-AN-LFD4 had an overall concordance of 95.3% (Additional file 1: Table S9) with a sensitivity of 74.9% and specificity of 99.2% (Table 1, Fig. 3). Sensitivity was higher (97.6%) at higher viral concentration (Additional file 1: Table S10). Comparison of sensitivity of SS with LFDs in studies SS-AN-LFD4 (Innova LFD at home, anterior nares) SS-NT-LFD1 (Innova, nose and throat), SS-NT-LFD3 (Innova, nose and throat), and SS-AN-LFD1 (Orient Gene, anterior nares) suggested that the at-home anterior nares testing approach used in SS-AN-LFD4 had better sensitivity (both overall and at high viral concentration) and more likely to correctly predict a positive result than the approaches used in studies SS-NT-LFD1, SS-NT-LFD3, and SS-AN-LFD1 (Table 3).

Table 3 Pairwise comparison of the sensitivity of self-swabbing of anterior nares vs nose and throat

Overall, these findings demonstrated that self-swabbing at home was as effective as swabbing at dedicated test sites for testing with LFDs.

Discussion

Throughout the SARS-CoV-2 pandemic, NHSTT aimed to provide the most effective testing approach on a national scale, by increasing the availability of SARS-CoV-2 tests and the speed of the testing process [1]. From the beginning of the program, it was recognized that the more invasive the swabbing approach, the more of a challenge sample collection would be in increasing test-seeking behavior and test accessibility. For the majority of the pandemic when symptomatic PCR testing was used, a negative test result had a direct impact on the individual, as the individual and their contacts did not need to continue self-isolating; therefore, identifying alternative approaches for those who could not complete a nose and throat swab was also an important element of accessibility. For regular twice-a-week asymptomatic LFD testing, factors likely to impact ongoing compliance were also key [37, 38]. Consistent user feedback, as demonstrated in the service evaluations and survey research reported here, indicated that there was a preference for less invasive approaches such as nose-only swabbing instead of nose and throat swabbing. In particular, participants reported operational difficulties with throat swabbing such as identifying the tonsils, gagging and sickness, and incidents of broken swabs. The observational data from the service evaluations supported the preference for nose over throat swabbing. The feedback on nose-only swabbing showed that the location and depth of swabbing impacted user experience. A preference was reported for shallow nose (anterior nares) swabbing, with feedback from study SS-MT-LFD1 reporting some negative experiences with (deeper nose) mid-turbinate.

Demonstration of the acceptable use of nose-only swabbing was expected to have several benefits, including increased tolerability of swabbing and increased test-seeking behavior, decreased risk of harm associated with throat swabbing, and alignment of instructions for users of different LFDs. However, without understanding the effect of the different swabbing methods, informed decision-making on UK policy could not be implemented. To this end, the program of service evaluation studies performed by NHSTT during the SARS-CoV-2 pandemic included those that evaluated the effectiveness and user acceptance of new testing and swabbing approaches (e.g., nose-only swabbing to the mid-turbinate level or anterior nares). The results subsequently reported here helped to inform decision-making and UK policy regarding SARS-CoV-2 testing approaches and techniques and supported the decision for nose-only LFD swabbing to be included in the COVID-19 National Testing Programme [6, 29].

Importantly, validation of different swabbing approaches needed to be considered against the potential impact of reduced sensitivity and false negatives, as well as self-isolation requirements [39]. SARS-CoV-2 testing in the UK was initially rolled out through a population-specific service delivery model comprising PCR testing with nose and throat swabbing for symptomatic individuals at dedicated test sites [5, 11]. Findings from Study SS-AN-PCR1 demonstrated a reduced sensitivity (of ~ 10%) with PCR analysis for nose-only swabbing compared with nose and throat swabbing, which prevented its widespread use in the context of symptomatic PCR testing. However, nose-only swabbing with PCR was still deemed adequate for public health testing as a viable alternative for those who could not perform nose and throat swabbing [40, 41]. For example, the use of nose-only swabbing with PCR could allow key workers who were required to self-isolate, but who could not perform nose and throat swabbing, to test and return to work if they had a negative PCR result when they were otherwise required to remain in self-isolation [42, 43].

From April 2021, LFDs were made freely and universally available in the UK as part of mass community testing, and asymptomatic individuals were encouraged to test twice weekly to identify SARS-CoV-2-positive cases and further reduce community transmission [16]. Findings from the initial AS-NT-LFD1 and SS-NT-LFD1 studies supported that swabbing of the nose and throat with LFDs provided sufficient diagnostic sensitivity (at least 50%) in a real-world setting for use in the NHSTT program [17,18,19,20, 33,34,35]. A sensitivity threshold of 50% was initially deemed acceptable for the use of LFDs and formed the basis for the original authorizations granted by the Medicines and Healthcare Products Regulatory Agency for the first self-test LFD [23]. Sensitivity was higher for high viral concentration, which is associated with increased infectiousness [17, 18, 27, 44], which allowed the rollout of population-wide asymptomatic testing with LFDs. Several subsequent service evaluations (studies AS-AN-LFD1, AS-AN-LFD2, SS-AN-LFD1 to SS-AN-LFD4, and SS-MT-LFD1) demonstrated a similar sensitivity of LFD tests to identify SARS-CoV-2 when using nose only (anterior nares [shallow nose] and mid-turbinate [deeper nose]) swabbing compared with nose and throat, supporting the use of nose only swabbing as a viable alternative and further increasing access to testing. The lower sensitivities of LFDs for identifying SARS-CoV-2 compared with PCR are mitigated by the different testing purposes for which they were predominantly used, with a likely increase in testing uptake and repeated testing [16, 45]. Modeling work has shown that repeated testing can contribute to a reduction in transmission [15, 46].

Across the service evaluation studies, using PCR and LFD, nose-only swabbing showed a high sensitivity for identifying SARS-CoV-2 among individuals with high viral concentrations (> 1 M copies/mL). The ability of nose-only swabbing to identify individuals with a high viral concentration is valuable because such individuals are most likely to be infectious [36, 41, 47]. For example, one study demonstrated that less than half of PCR-positive asymptomatic individuals were shedding SARS-CoV-2 whereas LFDs had more than 80% sensitivity in detecting those shedding SARS-CoV-2 [48].

While there was a reported preference for nose-only swabbing compared with nose and throat swabbing, the post-hoc analysis conducted by NHSTT did highlight that some uncertainty exists among test users regarding how far to insert the swab within the nostrils. While this is not a major concern in terms of barriers to testing accessibility, further research or supportive guidance on this issue may be appropriate in the future. Furthermore, the impact of the user swabbing experience on adherence and uptake was not assessed in the service evaluations. However, it is plausible to assume that the reported discomfort of throat swabbing may act as a deterrent to testing and that less invasive techniques, such as nose-only swabbing, may encourage test-seeking behavior.

It is important to note that the service evaluations were not designed to directly compare differences in diagnostic performance between swabbing approaches, but rather performance with respect to a chosen reference swabbing approach, and thus the cross-study comparisons presented here were exploratory and post-hoc. Changes in confounding variables might have an impact on the results and their robustness. In particular, the series of service evaluations assessed various LFDs and swabbing approaches and were conducted at different stages of the pandemic, and thus potential confounding with other variables (e.g., variants and vaccination) might exist in the comparison across service evaluation studies. Service evaluations did not compare directly to benchmarked standards for different use cases, such as symptomatic and asymptomatic mass testing, though evaluations of the testing program suggest performance was adequate to have a substantial public health impact [4]. Participation used a convenience sampling method of only asking site attendees to be tested, and thus study populations were not necessarily representative of the general population, but instead of those able and willing to attend regional and local testing sites. A similar caution is warranted in the interpretation of the user survey data and their representativeness, these were developed as part of an operational response and conformed with best practice for quality feedback. Finally, the data collection period ended in December 2021, prior to the dominance of the Omicron variants in the UK. Given the differences between the Omicron variants and earlier strains, the findings regarding the diagnostic accuracy of various swabbing techniques may not fully apply to the current situation.

In response to future pandemics, there need to be preliminary investigations assessing test and sample development to maximize testing accessibility and usability among users and encourage test-seeking behavior [49,50,51,52,53]. For any pandemic requiring mass community testing, the invasiveness of a swabbing approach is always likely to be a barrier to testing [53]. There may be instances where the sample collection and testing approach interfere with the accessibility to testing, and there may be other scenarios where the approach is less relevant to the individual but may affect the overarching intervention [52, 53]. The work reported here in the context of the SARS-CoV-2 pandemic demonstrates that mass community testing can be conducted using less invasive techniques. Although further validation is required, it is not unreasonable to assume that the results can be generalized for future scenarios and would be a basis for implementation, at least in the early stages while further validation is undertaken. Importantly, we demonstrated that testing performance is not affected by whether individuals test at home or at a testing site which further improves accessibility to testing [4, 33, 54].

Conclusions

User experience analyses performed in the UK during the SARS-CoV-2 pandemic supported a preference for nose-only swabbing compared with nose and throat swabbing. The findings from the service evaluation studies performed by NHSTT demonstrated the acceptable diagnostic performance of nose-only swabbing for LFDs in a real-world setting and supported the use of nose-only swabbing (for LFDs) as a viable option for use in the UK national testing initiative. The results of these studies informed and supported policy decisions regarding SARS-CoV-2 testing in the UK during the early phase of the SARS-CoV-2 pandemic and emphasized the importance of less invasive testing approaches to maximize testing accessibility in future pandemic responses.

Data availability

The datasets generated and/or analyzed during the current study are not publicly available as they were collected as part of a public health response and are considered sensitive. Access to anonymized data will be considered by the corresponding author on reasonable request.

Abbreviations

AN:

Anterior nares

AS:

Assisted swabbing

CI:

Confidence interval

Ct:

Cycle threshold

DHSC:

Department of Health and Social Care

FN:

False negative

FP:

False positive

HRA:

Health Research Authority

LFD:

Lateral flow device

MT:

Mid-turbinate

N/A:

Not applicable

NHSTT:

National Health Service Test and Trace

NT:

Nose and throat

PCR:

Polymerase chain reaction

PHE REGG:

Public Health England’s Research Ethics and Governance Group

qRT-PCR:

Quantitative reverse transcription-polymerase chain reaction

SEM:

Standard error

SS:

Self-swabbing

TN:

True negative

TP:

True positive

UKHSA:

UK Health Security Agency

VOC:

Variant of concern

VOTC:

Voice of the Customer

References

  1. Government launches NHS Test and Trace service. https://www.gov.uk/government/news/government-launches-nhs-test-and-trace-service.

  2. NHS to ramp up coronavirus testing labs. https://www.england.nhs.uk/2020/03/nhs-to-ramp-up-coronavirus-testing-labs/.

  3. Testing for coronavirus (COVID-19) will increase to 25,000 a day. https://www.gov.uk/government/news/testing-for-coronavirus-covid-19-will-increase-to-25-000-a-day.

  4. Evaluation of the national COVID-19 testing programme in England between October 2020 and March 2022. https://assets.ey.com/content/dam/ey-sites/ey-com/en_uk/resources/evaluation-of-covid-19-testing-in-england-eoha-conference-edition.pdf.

  5. Everyone in the United Kingdom with symptoms now eligible for coronavirus tests. https://www.gov.uk/government/news/everyone-in-the-united-kingdom-with-symptoms-now-eligible-for-coronavirus-tests.

  6. COVID-19: testing initiative evaluation programme. https://www.gov.uk/government/collections/covid-19-testing-initiative-evaluation-programme.

  7. Engstrom-Melnyk J, Rodriguez PL, Peraud O, Hein RC: Chapter 5 - clinical applications of quantitative real-time PCR in virology. In: Methods in microbiology. Volume 42, edn. Edited by Sails A, Tang Y-W: Academic Press; 2015: 161–197.

  8. Corman VM, Landt O, Kaiser M, Molenkamp R, Meijer A, Chu DK, Bleicker T, Brunink S, Schneider J, Schmidt ML, et al. Detection of 2019 novel coronavirus (2019-nCoV) by real-time RT-PCR. Euro Surveill. 2020;25(3):2000045.

    Article  PubMed  PubMed Central  Google Scholar 

  9. Yu F, Yan L, Wang N, Yang S, Wang L, Tang Y, Gao G, Wang S, Ma C, Xie R, et al. Quantitative detection and viral load analysis of SARS-CoV-2 in infected patients. Clin Infect Dis. 2020;71(15):793–8.

    Article  CAS  PubMed  Google Scholar 

  10. Kwok HF. The significance of advanced COVID-19 diagnostic testing in pandemic control measures. Int J Biol Sci. 2022;18(12):4610–7.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  11. Coronavirus (COVID-19): scaling up testing programmes. https://www.gov.uk/government/publications/coronavirus-covid-19-scaling-up-testing-programmes.

  12. PCR testing for SARS-CoV-2 during the COVID-19 pandemic. https://www.gov.uk/government/publications/pcr-testing-for-sars-cov-2-during-the-covid-19-pandemic.

  13. Antigen-detection in the diagnosis of SARS-CoV-2 infection - Interim Guidance. https://www.who.int/publications/i/item/antigen-detection-in-the-diagnosis-of-sars-cov-2infection-using-rapid-immunoassays.

  14. Wertenauer C, Pfeifer C, Roskos M, Marz W. Rapid antigen tests for SARS-CoV-2-a synopsis of the medical evidence. Diagn Microbiol Infect Dis. 2023;107(2):116027.

    Article  CAS  PubMed  Google Scholar 

  15. Quilty BJ, Clifford S, Hellewell J, Russell TW, Kucharski AJ, Flasche S, Edmunds WJ. Centre for the mathematical modelling of infectious diseases C-wg: quarantine and testing strategies in contact tracing for SARS-CoV-2: a modelling study. Lancet Public Health. 2021;6(3):e175–83.

    Article  PubMed  PubMed Central  Google Scholar 

  16. COVID-19: general public testing behaviours. https://www.gov.uk/government/publications/lfd-tests-how-and-why-they-were-used-during-the-pandemic/covid-19-general-public-testing-behaviours.

  17. Peto T. Team UC-LFO: COVID-19: Rapid antigen detection for SARS-CoV-2 by lateral flow assay: a national systematic evaluation of sensitivity and specificity for mass-testing. EClinicalMedicine. 2021;36:100924.

  18. Futschik ME, Johnson S, Turek E, Chapman D, Carr S, Thorlu-Bangura Z, Klapper PE, Sudhanva M, Dodgson A, Cole-Hamilton JR, et al. Rapid antigen testing for SARS-CoV-2 by lateral flow assay: a field evaluation of self- and professional testing at UK community testing sites. J Clin Virol. 2024;171:105654.

    Article  CAS  PubMed  Google Scholar 

  19. Garcia-Finana M, Hughes DM, Cheyne CP, Burnside G, Stockbridge M, Fowler TA, Fowler VL, Wilcox MH, Semple MG, Buchan I. Performance of the Innova SARS-CoV-2 antigen rapid lateral flow test in the Liverpool asymptomatic testing pilot: population based cohort study. BMJ. 2021;374:n1637.

    Article  PubMed  Google Scholar 

  20. Asymptomatic testing for SARSCoV-2 using antigen-detecting lateral flow devices. https://assets.publishing.service.gov.uk/media/60e56b3ee90e0764c6eb3866/asymptomatic-testing-for-SARS-CoV-2-using-antigen-detecting-lateral-flow-devices-evidence-from-performance-data-Oct-2020-to-May-2021.pdf.

  21. Wallace S, Hall V, Charlett A, Kirwan PD, Cole M, Gillson N, Atti A, Timeyin J, Foulkes S, Taylor-Kerr A, et al. Impact of prior SARS-CoV-2 infection and COVID-19 vaccination on the subsequent incidence of COVID-19: a multicentre prospective cohort study among UK healthcare workers - the SIREN (Sarscov2 Immunity & REinfection EvaluatioN) study protocol. BMJ Open. 2022;12(6):e054336.

    Article  PubMed  Google Scholar 

  22. Mass testing for COVID-19 using lateral flow tests. https://post.parliament.uk/mass-testing-for-covid-19-using-lateral-flow-tests/.

  23. Transparency data: COVID-19 self-test LFD exceptional use application to MHRA. https://www.gov.uk/government/publications/covid-19-self-test-lfd-exceptional-use-application-to-mhra.

  24. Ramdas K, Darzi A, Jain S. “Test, re-test, re-test”: using inaccurate tests to greatly increase the accuracy of COVID-19 testing. Nat Med. 2020;26(6):810–1.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  25. Raffle AE. Screening the healthy population for covid-19 is of unknown value, but is being introduced nationwide. BMJ. 2020;371:m4438. https://doiorg.publicaciones.saludcastillayleon.es/10.1136/bmj.m4438.

    Article  PubMed  Google Scholar 

  26. Larremore DB, Wilder B, Lester E, Shehata S, Burke JM, Hay JA, Tambe M, Mina MJ, Parker R. Test sensitivity is secondary to frequency and turnaround time for COVID-19 screening. Sci Adv. 2021;7(1):eabd5393.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  27. Lee LYW, Rozmanowski S, Pang M, Charlett A, Anderson C, Hughes GJ, Barnard M, Peto L, Vipond R, Sienkiewicz A, et al. Severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) infectivity by viral load, S gene variants and demographic factors, and the utility of lateral flow devices to prevent transmission. Clin Infect Dis. 2022;74(3):407–15.

    Article  CAS  PubMed  Google Scholar 

  28. Mina MJ, Parker R, Larremore DB. Rethinking Covid-19 test sensitivity - a strategy for containment. N Engl J Med. 2020;383(22):e120.

    Article  CAS  PubMed  Google Scholar 

  29. Scientific evidence supporting the government response to coronavirus (COVID-19). https://www.gov.uk/government/collections/scientific-evidence-supporting-the-government-response-to-coronavirus-covid-19.

  30. Research and analysis - investigation of SARS-CoV-2 variants: technical briefings. https://www.gov.uk/government/publications/investigation-of-sars-cov-2-variants-technical-briefings#full-publication-update-history.

  31. Guidance: COVID-19 Winter Plan. https://www.gov.uk/government/publications/covid-19-winter-plan/covid-19-winter-plan.

  32. Oral statement to Parliament: PM statement on COVID-19 Winter Plan: 23 November 2020. https://www.gov.uk/government/speeches/pm-statement-on-covid-19-winter-plan-23-november-2020.

  33. Eyre DW, Futschik M, Tunkel S, Wei J, Cole-Hamilton J, Saquib R, Germanacos N, Dodgson AR, Klapper PE, Sudhanva M, et al. Performance of antigen lateral flow devices in the UK during the alpha, delta, and omicron waves of the SARS-CoV-2 pandemic: a diagnostic and observational study. Lancet Infect Dis. 2023;23(8):922–32.

    Article  CAS  PubMed  Google Scholar 

  34. Testing for SARS-CoV-2 using antigen detecting lateral flow devices. Evidence from performance assessment and evaluation for the period February 2021 to March 2022. https://assets.publishing.service.gov.uk/media/6416dc91d3bf7f79d9675d96/SARS-CoV-2-testing-using-LFD-evidence-from-performance-data-part-2.pdf.

  35. Evaluation of lateral flow device performance within the National Testing Programme: Date of reporting period: 8 November 2020 to 21 March 2022.https://assets.publishing.service.gov.uk/media/638a1ff18fa8f569f4902038/LFD_Performance_within_the_NTP_Report.pdf.

  36. Research and analysis: lateral flow device (LFD) performance data. https://www.gov.uk/government/publications/lateral-flow-device-performance-data.

  37. Alhakmi F, Alhakmi D, Mimoglu E. Exploring COVID-19 lateral flow testing engagement and compliance in selected Imperial College Healthcare Trust wards. Clin Med (Lond). 2022;22(Suppl 4):78–9.

    Article  PubMed  Google Scholar 

  38. Love NK, Ready DR, Turner C, Yardley L, Rubin GJ, Hopkins S, Oliver I. The acceptability of testing contacts of confirmed COVID-19 cases using serial, self-administered lateral flow devices as an alternative to self-isolation. J Med Microbiol. 2022;71(8):001567.

    Article  Google Scholar 

  39. Impact of false-positives and false-negatives in the UK’s COVID-19 RT-PCR testing programme, 3 June 2020. https://www.gov.uk/government/publications/gos-impact-of-false-positives-and-negatives-3-june-2020/impact-of-false-positives-and-false-negatives-in-the-uks-covid-19-rt-pcr-testing-programme-3-june-2020.

  40. Guidance: Technologies Validation Group: using tests to detect COVID-19. https://www.gov.uk/guidance/technologies-validation-group-using-tests-to-detect-covid-19.

  41. Fowler T, Blandford E, Chapman D, Futschik ME, Kulasegaran-Shylini R, Tunkel S, Lewis C, Fellows A, Sheppard E, McCabe L, et al. Comparative evaluation of swabbing sites for Omicron variant detection in PCR testing. Diagn Microbiol Infect Dis. 2025;111(1):116577.

    Article  CAS  PubMed  Google Scholar 

  42. Guidance: Stay at home: guidance for households with possible or confirmed coronavirus (COVID-19) infection [withdrawn]. https://www.gov.uk/government/publications/covid-19-stay-at-home-guidance/stay-at-home-guidance-for-households-with-possible-coronavirus-covid-19-infection.

  43. Guidance: COVID-19: management of staff and exposed patients or residents in health and social care settings [withdrawn]. https://www.gov.uk/government/publications/covid-19-management-of-exposed-healthcare-workers-and-patients-in-hospital-settings/covid-19-management-of-exposed-healthcare-workers-and-patients-in-hospital-settings.

  44. Bhavnani D, James ER, Johnson KE, Beaudenon-Huibregtse S, Chang P, Rathouz PJ, Weldon M, Matouschek A, Young AE. SARS-CoV-2 viral load is associated with risk of transmission to household and community contacts. BMC Infect Dis. 2022;22(1):672.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  45. Fearon E, Buchan IE, Das R, Davis EL, Fyles M, Hall I, Hollingsworth TD, House T, Jay C, Medley GF, et al. SARS-CoV-2 antigen testing: weighing the false positives against the costs of failing to control transmission. Lancet Respir Med. 2021;9(7):685–7.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  46. Whitfield CA. University of Manchester C-MG, Hall I: Modelling the impact of repeat asymptomatic testing policies for staff on SARS-CoV-2 transmission potential. J Theor Biol. 2023;557:111335.

    Article  CAS  PubMed  Google Scholar 

  47. Evidence summary for lateral flow devices (LFD) in relation to care homes. https://www.gov.uk/government/publications/evidence-on-the-accuracy-of-lateral-flow-device-testing/evidence-summary-for-lateral-flow-devices-lfd-in-relation-to-care-homes.

  48. Petersen I, Crozier A, Buchan I, Mina MJ, Bartlett JW. Recalibrating SARS-CoV-2 antigen rapid lateral flow test relative sensitivity from validation studies to absolute sensitivity for indicating individuals shedding transmissible virus. Clin Epidemiol. 2021;13:935–40.

    Article  PubMed  PubMed Central  Google Scholar 

  49. Smith LE, Potts HW, Amlot R, Fear NT, Michie S, Rubin GJ. Who is engaging with lateral flow testing for COVID-19 in the UK? The COVID-19 Rapid Survey of Adherence to Interventions and Responses (CORSAIR) study. BMJ Open. 2022;12(2):e058060.

    Article  PubMed  Google Scholar 

  50. Green K, Micocci M, Hicks T, Winter A, Martin JE, Shinkins B, Shaw L, Price C, Davies K, Allen JA. Perceived feasibility, facilitators and barriers to incorporating point-of-care testing for SARS-CoV-2 into emergency medical services by ambulance service staff: a survey-based approach. BMJ Open. 2022;12(11):e064038.

    Article  PubMed  Google Scholar 

  51. Green MA, Garcia-Finana M, Barr B, Burnside G, Cheyne CP, Hughes D, Ashton M, Sheard S, Buchan IE. Evaluating social and spatial inequalities of large scale rapid lateral flow SARS-CoV-2 antigen testing in COVID-19 management: an observational study of Liverpool, UK (November 2020 to January 2021). Lancet Reg Health Eur. 2021;6:100107.

    Article  PubMed  PubMed Central  Google Scholar 

  52. Plantamura J, Bousquet A, Otto MP, Bigaillon C, Legland AM, Delacour H, Vest P, Astier H, Valero E, Bylicki O, et al. Performances, feasibility and acceptability of nasopharyngeal swab, saliva and oral-self sampling swab for the detection of severe acute respiratory syndrome coronavirus 2. Eur J Clin Microbiol Infect Dis. 2021;40(10):2191–8.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  53. Theel ES, Kirby JE, Pollock NR. Testing for SARS-CoV-2: lessons learned and current use cases. Clin Microbiol Rev. 2024;37(2):e0007223.

    Article  PubMed  Google Scholar 

  54. Self-testing and self-care. https://www.finddx.org/what-we-do/cross-cutting-workstreams/self-testing/.

Download references

Acknowledgements

Medical writing support for the development of this manuscript, under the direction of the authors, was provided by Ewan Smith, PhD of Ashfield MedComms, an Inizio company, and funded by the UKHSA.

Funding

The service evaluation program was funded by UKHSA’s predecessor organization NHS Test and Trace (part of the UK Department of Health and Social Care) and UK Government Department of Health and Social Care.

Author information

Authors and Affiliations

Authors

Contributions

TF, MF, RK-S, EB, DC conceived and designed the research. MF, RK-S, EB, SH, DC, ET, SA, VP, HF, LC, MK, AD, PK, MS, TP, ST were responsible for data acquisition, analysis & interpretation: All authors reviewed and edited the manuscript.

Corresponding author

Correspondence to Tom Fowler.

Ethics declarations

Ethics approval and consent to participate

Within the context of the pandemic public health response and roll-out of testing interventions, after review using the UK Health Research Authority (HRA) tool (https://hra-decisiontools.org.uk/research/) and further discussions with HRA, it was determined that these evaluations (described in this report) did not require HRA ethics approval. After an initial period, it was determined to gain Public Health England’s Research Ethics and Governance Group (PHE REGG) approval (then separate to NHSTT) for service evaluations for subsequent studies to ensure further external scrutiny and assurance on this approach. Approval was obtained for an umbrella framework (UKHSA framework for evaluating the safety and effectiveness of devices and new testing methods for diagnosing COVID-19 and other infectious diseases (ISRCTN-ISRCTN50943319)) and associated participant-facing materials for the prospective data collection elements of Service Evaluation and Ongoing Evaluation. This was reviewed and approved under IRAS project ID: 317913 and REC reference-22/SC/0340. All study participants received routine care through receipt of an individual diagnostic swab test and result. Participants gave verbal documented consent to participate in the evaluation. The service evaluations conformed to the ethical principles of the Declaration of Helsinki.

Consent for publication

Not applicable.

Competing interests

David Chapman is an employee of Deloitte MCS Ltd. All remaining authors declare no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

44263_2024_121_MOESM1_ESM.docx

Additional file 1: Supplementary methods, Supplementary results, Table S1: Study ID of service evaluations, Table S2: Numbers of void samples and other omitted samples in the service evaluations, Table S3: SS-AN-PCR1: Concordance matrix of nose and throat swabbing versus nose only (anterior nares) swabbing, Table S4: SS-AN-PCR1 (anterior nares): quantitative data analysis by Ct. Table S5: Concordance matrix of LFD and PCR results in assisted-swabbing studies: AS-NT-LFD1 (nose and throat), AS-AN-LFD1 (anterior nares), and AS-AN-LFD2 (anterior nares). Table S6: Sensitivity of assisted swabbing using LFDs overall and by viral concentration, compared with assisted swabbing of the nose and throat with PCR, in studies AS-NT-LFD1 (nose and throat), AS-AN-LFD1 (anterior nares), and AS-AN-LFD2 (anterior nares). Table S7: Concordance matrix of LFD and PCR results in Studies SS-NT-LFD1 (nose and throat), SS-NT-LFD2 (nose and throat), SS-AN-LFD1 (anterior nares), SS-AN-LFD2 (anterior nares), and SS-MT-LFD1 (mid-turbinate level). Table S8: Sensitivity of self-swabbing using LFDs overall and by viral concentration, compared with self-swabbing of the nose and throat with PCR, in Studies SS-NT-LFD1 (nose and throat), SS-AN-LFD1 (anterior nares), SS-AN-LFD2 (anterior nares), SS-NT-LFD2 (nose and throat), and SS-MT-LFD1 (mid-turbinate level). Table S9: Concordance matrix of LFD and PCR results in Studies SS-NT-LFD3 (nose and throat), SS-AN-LFD3 (anterior nares), and SS-AN-LFD4 (anterior nares). Table S10: Sensitivity of at-home self-swabbing using LFDs overall and by viral concentration, compared with self-swabbing of the nose and throat with PCR, in Studies SS-NT-LFD3 (nose and throat), SS-AN-LFD3 (anterior nares), and SS-AN-LFD4 (anterior nares). Fig. S1: User experience findings from the LFD Product Research Team. A Ease of use with nose and throat swabbing (N=54,415). B Recorded incidents of broken swabs with nose and throat swabbing (PCR and LFD). Fig. S2 Word cloud summary of detractor descriptions for SARS-CoV-2 swabbing.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Futschik, M.E., Kulasegaran-Shylini, R., Blandford, E. et al. Effectiveness and user experience of nose and throat swabbing techniques for SARS-CoV-2 detection: results from the UK COVID-19 National Testing Programme. BMC Glob. Public Health 3, 5 (2025). https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s44263-024-00121-x

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s44263-024-00121-x

Keywords